Biowave 3 and 3+ UV-Visible Life Science Spectrophotometers
Biowave 3 and 3+ UV-Visible Life Science Spectrophotometers
The Biowave 3 offers all the benefits and capabilities of a research grade Life Science spectrophotometer with the robust design and reasonable cost that also makes it perfect for students. The Biowave 3+ is a higher resolution 3nm bandwidth version.
Built in Life Science Methods
Predefined life science methods for nucleic acid, protein and microbial culture quantification and analysis, combined with the rapid scanning, kinetics and concentration capabilities, and their compatibility with disposable low volume UV cuvettes have made them a common sight among educational and industrial laboratories alike.
Feature:
- Split Beam (RBC) UV/Vis Spectrophotometer 190-1100nm (3nm/5nm bandwidth)
- Pre-programmed methods for DNA, RNA, Oligos, proteins and cell density
- Nucleic acid scans for purity checking
- Low volume capability with Microvolume cuvettes
- Novel optics with a Xenon source for long lamp lifetime and low running costs
- Lightweight and stackable for storage
Flexible Data Export Options
Results may be downloaded to a USB flash drive or printed to an optional integrated graphical printer for a permanent record. Alternatively, results can be exported via a cable connection to a suitable PC running Print Via Computer (PVC) software or wireless accessory (optional).
Comprehensive IQ/OQ Support for your Biowave instrument:
Comprehensive IQ/OQ documentation and service support is available for our Biowave instruments allowing the qualification of wavelength accuracy, wavelength repeatability, photometric accuracy, stray light, resolution, noise, and drift.
If your lab requires IQ (Installation Qualification) and OQ (Operational Qualification), contact us to learn more.
Technical Specifications | ||
Specification | Data | |
Wavelength Range | 190 to 1100 nm | |
Monochromator | Flat grating | |
Wavelength Calibration | Automatic upon switch on | |
Beam Height | 15 mm | |
Spectral Bandwidth | 5 nm or 3 nm for ‘+’ models | |
Wavelength Accuracy | ±2 nm | |
Wavelength Reproducibility | ±1 nm | |
Light Source | Xenon flash lamp | |
Detector | Twin CMOS array | |
Photometric Range | 0.0 to 2.500 A, 0.3 to 100 %T | |
Photometric Linearity | ±1.3 % or ±0.008 A whichever is greater at 546 nm | |
Photometric Reproducibility | ±0.002 A to 0.5 A at 546 nm | |
Stray Light | <0.5 %T 340 nm | |
Stability | ±0.01 A/h at 340 nm | |
Noise | ±0.005 peak to peak ±0.002 RMS | |
Digital Output | USB flash drive, PC via PVC software, Optional Wireless | |
Data Export | USB flash drive: .tsv, native PVC format PC via PVC: .csv, .emf, .xlsx, .xls, .rtf, .tsv, native PVC format | |
Method Storage | 90 with PIN number protection | |
Graphical Display | Yes, zoom and track function | |
Sample ID | Yes | |
Languages | English, German, French, Spanish, Italian, Japanese, Chinese | |
Dimensions | 260 × 390 × 100 mm | |
Weight 3.00 kg | 3.00 kg |
Related Documents | ||
Biowave 3 and 3+ UV-Visible Life Science Spectrophotometers.pdf | DOWNLOAD BROCHURE |
Document contains all procedures required to ensure the installation and qualification of the system according to the manufacturer’s specification.